АТ1316A Whole Body Counter

AT1316A Whole Body Counter (WBC) is designed for express-monitoring and measuring of gamma-emitting radionuclides in human lungs ("Sitting straight" geometry).
FEATURES
 
  • Stabilized spectrometric path
  • Spectrometric and radiometric measurement modes
  • Total activity threshold excess monitoring for 51Cr, 54Mn, 58Co, 59Fe, 65Zn, 95Nb, 100mAg, 103Ru, 124Sb, 141Ce and 144Ce in lungs
  • Efficient and highly sensitive algorithm for instrument spectra processing utilizing matrix method for activity estimation
  • Radionuclide identification in spectrometric mode
  • Flexible software control of spectrometer functions, generation of database and report based on measurement results
  • Fixed chair geometry
  • Compact design
  • Prompt accommodation to background conditions using operational background generating option
  • Can be installed into a van as part of mobile radiation monitoring laboratory
  • USB Counter-to-PC connection
APPLICATIONS
 
Citizens and staff individual dosimetric monitoring of internal exposure:
  • Citizens and staff monitoring during and after radiation accidents
  • Factory and office workers monitoring, involved into radioactive material production or use
  • Nuclear Power Plant staff monitoring
OPERATING PRINCIPLE
 
WBC operating principle is based on detection of incorporated gamma-emitting radionuclides activity in human lungs with spectrometric detection unit and processing of spectrometer measurement information with hardware-software instruments to define radiometric characteristic of internal contamination considering anthropometric characteristic of target person.

Measurement of a Person Joined version of AT1322 and AT1316/AT1316A

* The Customer shall purchase a PC and a printer individually or use already available equipment (Minimum technical requirements to PC shall be agreed with the Manufacturer).

Features



Measurement result display
Detector Scintillator, NaI(Tl) Ø150x100 mm
Registered gamma radiation energy range 50 keV...2 MeV
Minimum measured activity of 60Cs in adult human lungs for 3 min. measurement interval 60 Bq
60Co radionuclide activity measurement range in adult human lungs 40...1·105 Bq
Measurement geometry "Sitting straight”
Intrinsic relative error of activity measurement ±20%
Number of ADC channels 512
Integral nonlinearity ±1% max.
Typical resolution at 662 keV (137Cs) 9%
Measurement instability during continuous service 3% max.
Express-monitoring productivity 15 person/h
Operation mode setup time 10 min
Continuous run time 24 h
Working temperature range 10°C...35°C
Relative air humidity with air temperature ≤30°C without condensation ≤75%
Power supply 230 VAC , 50 Hz
Power consumption ≤200 VA
Weight 250 kg

AТ1316A Whole Body Counter  meets International standard requirements: IEC 61582:2004,
IEC 60601-1:2005, EN 50371:2002, Safety standard requirements: IEC 61010-1:1990,
EMC requirements: EN 55022:1998+A1:2000+A2:2003, EN 55024:1998+A1:2001+A2:2003,
IEC 61000-4-2-2001

Design and specifications are subject to change without notice

 